This is an old revision of the document!
1.Purpose of examination:
- AST estimation from serum NADH (without P-5’-P) Method.
2.Responsibility and Authority:
- Calibration: Technician
- Quality Control: Technician
- Routine operation: Technician
- Overall Monitoring: Quality Manager
3.Sample Details:
- Type of Sample: Serum,
- Type of container and additives: Plain without any additives
- Patient Preparation: As per Primary Sample Collection Manual sample_collection_manual
- Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours
- Handling and transport: As per Primary Sample collection manual
- Storage: 24 hours at 2-8° C
4.Required Equipment:
- Centrifuge, Auto-Pipette, Disposable Tips, Disposable sample cups, FullyAuto Chemistry Analyzer
5.Required reagents: R1 L-aspartic acid (103.860 g/L), lactate dehydrogenase (4.000 KU/L), and malate dehydrogenase (2.000 KU/L). R2 α-ketoglutaric acid (6.570 g/L).
6.Reagent Handling
- Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles
- Reagent Storage and stability
- Unopened reagent stable at 2-8°C until expiration date.
- On board System temperature reagent is stable for 30 days
- Instability or deterioration should be suspected if there are precipitates, visible signs of leakage or
- contamination, turbidity, or if calibration or controls do not meet the appropriate criteria.
7.Calibration Procedure:
- SGOT Calibrators-WATER BLANK
- Frequency:
- Reagent lot change
- QC out of range
- After service or maintenance
- Replacement in any parts of Instrument
Procedure:
- Start the equipment.WDI abbotte fully.docx
- Calibrators are ready to use.
- Put calibrator 15-20 minutes at room temperature
- Prior to use, mix the contents by inverting the vial. Do not shake the vial to prevent foam formation.
- Take a 150 µl calibrator solution in to separate aliquot.
- Go to the calibration and give the calibration order.
- Verify calibration with at least two levels of controls
- If control results fall outside acceptable ranges, root cause analysis or recalibration may be necessary.
8.Quality control Procedure: Name: Bio Rad Level 1 & 2 Frequency: As per Quality Control Procedure
Procedure for Reconstitution of IQC
- Reconstitution of QC material with 5 ml Distilled water by using calibrated fixed volume pipette.
- Leave to stand for 30 min in the dark place.
- Swirl gently several times during the reconstitution period to ensure that the contents are completely dissolved.
- Prior to use, mix the contents by inverting the vial. Do not shake the vial as the information of foam should be avoided. Ensure that no lyophilized material remains un-reconstituted.
- Prepare aliquots of 150 µl from the reconstituted QC material.
- Store these aliquots at -15° C to -20° C.
- Prior to use, make sure that aliquots should be at room temperature for at least 15 min.
Procedure to run IQC
- Press Control order
- Select Assay /Panel, to be run.
- Select the control/s and its level/s
- Give Carrier Number and Position number
- Press F3 / Add order
- Put respected carrier in RSH rack
- Check IQC results, in case outliers call residents.
9.Principle of the procedure used for examinations:
- AST present in the sample catalyzes the transfer of the amino group from L-aspartate to α-ketoglutarate, forming oxaloacetate and L-glutamate. Oxaloacetate in the presence of nicotinamide adenine dinucleotide (NADH) and malate dehydrogenase (MDH) is reduced to L-malate. In this reaction, NADH is oxidized to NAD+. The reaction is monitored by measuring the rate of decrease in absorbance at 340 nm due to the oxidation of NADH to NAD+.
10.Sample Preparation:
- Required SampleVolume: 150 µl of the sample
- Temperature: 37°
- Take 150-200µl of the sample from Primary tube to the secondary aliquot. Write the sample ID on the aliquot.
Procedure to run Patient sample
- Press Patient order
- Select Assay /Panel, to be run.
- Give Carrier Number and Position number
- Press F3 / Add order
- Put respected carrier in RSH rack
11.Performance Characteristics:
- Linearity: 5 to 4001 U/L
- The limit of detection (LOD): 4 U/L
- The limit of quantification(LOQ): 5 U/L
- Unit:U/L
Normal and critical ranges: Aspartate aminotransferase
male | <35 | U/L |
---|---|---|
Female | <31 | U/L |
12.Laboratory Clinical interpretation: AST is found to be elevated in alcohol-induced liver disease, viral hepatitis, cirrhosis and acute fulminant hepatic failure.
13.Interference and cross reaction: The Following analytes were tested up to the two levels indicated at SGOT concentrations of 30 U/L and 120 U/L, and found not to interfere:
- No interference from Bilirubin up to 30 mg/dL
- No interference from Hemoglobin up to 62 mg/dL
- No interference from Intralipid up to 550 mg/dL
- No interference from Sulfapyrdine up to 300 mg/dl
- No interference from Sufasalazine up to 300 mg/dl
- No interference from Temozolomyde up to20 mg/dl
14.Potential source of variation: Turn around time (TAT): Routine: 6.0 hours Urgent: 2.0 hours
15.Recording of observation:
- Software backup
- Machine raw data
16.Storage & Disposal of waste:Follow storage & discard procedure
17.Environmental & Safety control:
- For in vitro diagnostic use
- Do not use component before the expiration date
- Do not mix material from different kit lot numbers.
- Reagent 2 causes mild skin irritation.
18.References:
- Burtis CA,Ashwood ER,editors,Tietz Textbook of clinical chemistry.
- World health organization. Biosafety manual,3rd edition.
- Clinical and laboratory std institute . Protection of Laboratory workers from occupationally acquired infection; Approved guideline -4th edition.
- Burtis CA,Ashwood ER,editors,Tietz Textbook of clinical chemistry
- Kaplan LA,Pesce AJ editors. Clinical Chemistry Theory, Analysis, and Correlation, 3rd ed.