Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Next revision
Previous revision
tsh [2025/01/28 11:26] – created admintsh [2025/01/29 04:39] (current) admin
Line 1: Line 1:
 +|[[clinical_biochemistry_section|Home]]|[[clinical_biochemistry|]]|[[examination_procedures|]]|
  
-Purpose of examination: +                      TSH Examination Procedure 
 +**1.Purpose of examination:** 
 TSH assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of human Thyroid Stimulating Hormone (TSH) in human serum and plasma. TSH assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the quantitative determination of human Thyroid Stimulating Hormone (TSH) in human serum and plasma.
  
-Responsibility and Authority:  +**2.Responsibility and Authority:**  
-● Calibration: Technician  +  Calibration: Technician  
-● Quality Control: Technician  +  Quality Control: Technician  
-● Routine operation: Technician  +  Routine operation: Technician  
-● Overall Monitoring: Quality Manager+  Overall Monitoring: Quality Manager
  
- Sample Details:  +**3.Sample Details:**  
-● Type of Sample: Serum,  Plasma +  Type of Sample: Serum,  Plasma 
-● Type of container and additives: Plain without any additives  +  Type of container and additives: Plain without any additives  
-● Patient Preparation: As per Primary Sample Collection Manual {{Sample collection manual}} +  Patient Preparation: As per Primary Sample Collection Manual [[sample_collection_manual|]] 
-● Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours +  Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours 
-● Handling and transport: As per Primary Sample collection manual  +  Handling and transport: As per Primary Sample collection manual  
-● Storage: 24 hours at 2-8° C+  Storage: 24 hours at 2-8° C
  
-Required Equipment:  +**4.Required Equipment:**  
-Centrifuge, Auto-Pipette, Disposable Tips, Disposable sample cups, FullyAuto Chemistry Analyzer+Centrifuge, Auto-Pipette, Disposable Tips, Disposable sample cups, FullyAuto Chemistry Analyzer
  
-Required reagents: +**5.Required reagents:** 
-Microparticles +  Microparticles 
-Conjugate +  Conjugate 
-Multi Assay manual dilution +  Multi Assay manual dilution 
-Pre trigger solution +  Pre trigger solution 
-Trigger solution +  Trigger solution 
-Wash-buffer +  Wash-buffer 
  
-Reagent Handling +**6.Reagent Handling** 
- ● Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles +  Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles 
-Reagent Storage and stability +  Reagent Storage and stability 
-Unopened reagent stable at 15-30°C until expiration date. +  Unopened reagent stable at 15-30°C until expiration date. 
-On board System temperature reagent is stable for 28 days +  On board System temperature reagent is stable for 28 days 
-Instability or deterioration should be suspected if there are precipitates, visible signs of leakage or +  Instability or deterioration should be suspected if there are precipitates, visible signs of leakage or 
-contamination, turbidity, or if calibration or controls do not meet the appropriate criteria.+  contamination, turbidity, or if calibration or controls do not meet the appropriate criteria.
  
-Calibration Procedure: +**7.Calibration Procedure:** 
-ARCHITECT  TSH Calibrators +  ARCHITECT  TSH Calibrators 
-● Frequency: +  Frequency: 
-Reagent lot change +  Reagent lot change 
-QC out of range +  QC out of range 
-After service or maintenance +  After service or maintenance 
-Replacement in any parts of Instrument +  Replacement in any parts of Instrument 
-Procedure: +  Procedure: 
-Start the equipment.WDI abbotte fully.docx +  Start the equipment.WDI abbotte fully.docx 
-Calibrators are ready to use. +  Calibrators are ready to use. 
-Put calibrator 15-20 minutes at room temperature   +  Put calibrator 15-20 minutes at room temperature   
-Prior to use, mix the contents by inverting the vial. Do not shake the vial to prevent foam formation.  +  Prior to use, mix the contents by inverting the vial. Do not shake the vial to prevent foam formation.  
-Take a 150 µl calibrator solution in to separate aliquots. +  Take a 150 µl calibrator solution in to separate aliquots. 
-Go to the calibration and give the calibration order. +  Go to the calibration and give the calibration order. 
-Verify calibration with at least two levels of controls +  Verify calibration with at least two levels of controls 
-If control results fall outside acceptable ranges,root cause analysis or recalibration may be necessary.+  If control results fall outside acceptable ranges,root cause analysis or recalibration may be necessary.
  
-Quality control Procedure:+**8.Quality control Procedure:**
 Name: Bio Rad Level 1 ,2 & 3 Name: Bio Rad Level 1 ,2 & 3
 Frequency: As per Quality Control Procedure Frequency: As per Quality Control Procedure
-●Procedure for Reconstitution of IQC  
-Reconstitution of QC material with 5 ml Distilled water by using calibrated fixed volume pipette. 
-Leave to stand for 30 min in the dark place. 
-Swirl gently several times during the reconstitution period to ensure that the contents are completely dissolved. 
-Prior to use, mix the contents by inverting the vial. Do not shake the vial as the information of foam should be avoided. Ensure that no lyophilized material remains un-reconstituted. 
-Prepare aliquots of 150 µl from the reconstituted QC material. 
-Store these aliquots at -15° C to -20° C. 
-Prior to use, make sure that aliquots should be at room temperature for at least 15 min. 
-●Procedure to run IQC 
-Press Control order  
-Select Assay /Panel, to be run. 
-Select the control/s and its level/s 
-Give Carrier Number and Position number 
-Press F3 / Add order 
-Put respected carrier in RSH rack 
-Check IQC results, in case outliers call residents. 
  
-Principle of the procedure used for examinations:+**Procedure for Reconstitution of IQC**  
 +  * Reconstitution of QC material with 5 ml Distilled water by using calibrated fixed volume pipette. 
 +  * Leave to stand for 30 min in the dark place. 
 +  * Swirl gently several times during the reconstitution period to ensure that the contents are completely dissolved. 
 +  * Prior to use, mix the contents by inverting the vial. Do not shake the vial as the information of foam should be avoided. Ensure that no lyophilized material remains un-reconstituted. 
 +  * Prepare aliquots of 150 µl from the reconstituted QC material. 
 +  * Store these aliquots at -15° C to -20° C. 
 +  * Prior to use, make sure that aliquots should be at room temperature for at least 15 min. 
 +**Procedure to run IQC** 
 +  * Press Control order  
 +  * Select Assay /Panel, to be run. 
 +  * Select the control/s and its level/s 
 +  * Give Carrier Number and Position number 
 +  * Press F3 / Add order 
 +  * Put respected carrier in RSH rack 
 +  * Check IQC results, in case outliers call residents. 
 + 
 +**9.Principle of the procedure used for examinations:**
 The ARCHITECT TSH assay is an automated two-step immunoassay to determine the presence of Thyroid Stimulating Hormone (TSH) in human serum and plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex. The ARCHITECT TSH assay is an automated two-step immunoassay to determine the presence of Thyroid Stimulating Hormone (TSH) in human serum and plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex.
-1. Sample, anti-β TSH antibody coated paramagnetic microparticles and TSH Assay Diluent are combined. TSH present in the sample binds to the anti-TSH antibody coated microparticles. +  * Sample, anti-β TSH antibody coated paramagnetic microparticles and TSH Assay Diluent are combined. TSH present in the sample binds to the anti-TSH antibody coated microparticles. 
-2. After washing, anti-α TSH acridinium-labeled conjugate is added to create a reaction mixture. +  After washing, anti-α TSH acridinium-labeled conjugate is added to create a reaction mixture. 
-3. Following another wash cycle, Pre-Trigger and Trigger Solutions are added to the reaction mixture. +  Following another wash cycle, Pre-Trigger and Trigger Solutions are added to the reaction mixture. 
-4. The resulting chemiluminescent reaction is measured as relative light units (RLUs). There is a direct relationship between the amount of TSH in the sample and the RLUs detected by the +  The resulting chemiluminescent reaction is measured as relative light units (RLUs). There is a direct relationship between the amount of TSH in the sample and the RLUs detected by the ARCHITECT iSystem optics.
-ARCHITECT iSystem optics.+
  
- Sample Preparation: +**10.Sample Preparation:** 
- ● Required SampleVolume: 150 µl of the sample  +  Required SampleVolume: 150 µl of the sample  
- ● Temperature: 37° +  Temperature: 37° 
 Take 150-200µl of the sample from Primary tube to the secondary aliquot. Write the sample ID on the aliquot.  Take 150-200µl of the sample from Primary tube to the secondary aliquot. Write the sample ID on the aliquot. 
-Procedure to run Patient sample +**Procedure to run Patient sample** 
-Press Patient order  +  Press Patient order  
-Select Assay /Panel, to be run. +  Select Assay /Panel, to be run. 
-Give Carrier Number and Position number +  Give Carrier Number and Position number 
-Press F3 / Add order +  Press F3 / Add order 
-Put respected carrier in RSH rack+  Put respected carrier in RSH rack
  
-Performance Characteristics:  +**11.Performance Characteristics:**  
-● Linearity:  0.0038 μIU/mL to 100.0000 μIU/mL. +Linearity:  0.0038 μIU/mL to 100.0000 μIU/mL. 
-● The limit of detection (LOD) : ≤ 0.4 ng/dL. +The limit of detection (LOD) : ≤ 0.4 ng/dL. 
-● The limit of quantification(LOQ): 0.28 ng/dL +The limit of quantification(LOQ): 0.28 ng/dL 
-● Unit:ng/dL.+Unit:ng/dL.
  
  
- Normal and critical ranges:+**Normal and critical ranges:**
  
-Thyrotropin (thyroid-stimulating hormone) (TSH) +**Thyrotropin (thyroid-stimulating hormone) (TSH)** 
-Birth-4 d +^Birth-4 d^1.0-39.0^µIU/mL^ 
-1.0-39.0 +^2-20 wk^1.7-9.1^µIU/mL
-µIU/mL+^21 wk-20 y^0.7-6.4^µIU/mL^ 
 +^21-54 y^0.4-4.2^µIU/mL^ 
 +^55-87 y^0.5-8.9^µIU/m^L
  
- +**12.Laboratory Clinical interpretation:**
-2-20 wk +
-1.7-9.1 +
-µIU/mL +
- +
- +
-21 wk-20 y +
-0.7-6.4 +
-µIU/mL +
- +
- +
-21-54 y +
-0.4-4.2 +
-µIU/mL +
- +
- +
-55-87 y +
-0.5-8.9 +
-µIU/mL +
- +
- +
- Laboratory Clinical interpretation:+
 The common causes of High TSH Level are as follows: Hypothyroidism The common causes of LowTSH Level are as follows:Primary Hyperthyroidism The common causes of High TSH Level are as follows: Hypothyroidism The common causes of LowTSH Level are as follows:Primary Hyperthyroidism
  
- Interference and cross reaction:+**13.Interference and cross reaction:**
 TSH assay is designed to have a potential interference from hemoglobin, bilirubin, triglycerides and protein of ≤ 10% at the levels indicated below. TSH assay is designed to have a potential interference from hemoglobin, bilirubin, triglycerides and protein of ≤ 10% at the levels indicated below.
-Hemoglobin          ≤ 500 mg/dL +^Hemoglobin^≤ 500 mg/dL^ 
-Bilirubin               ≤ 20 mg/dL +^Bilirubin^≤ 20 mg/dL^ 
-Triglycerides         ≤ 3000 mg/dL +^Triglycerides^≤ 3000 mg/dL^ 
-Protein                  ≤ 2 g/dL and 12 g/dL+^Protein^≤ 2 g/dL and 12 g/dL^
  
-Potential source of variation: +**14.Potential source of variation:** 
 Turn around time (TAT): Turn around time (TAT):
 Routine: 6.0 hours Routine: 6.0 hours
 Urgent: 2.0 hours Urgent: 2.0 hours
  
- Recording of observation:+**15.Recording of observation**:
 Software backup  Software backup 
 Machine raw data Machine raw data
-Storage & Disposal of waste: Follow storage & discard procedure + 
-Environmental & Safety control:+**16.Storage & Disposal of waste:** Follow storage & discard procedure 
 + 
 +**17.Environmental & Safety control:**
 Contact with acids liberates very toxic gas. Dispose of contents / container in accordance with local regulations. Contact with acids liberates very toxic gas. Dispose of contents / container in accordance with local regulations.
  
-References: +**18.References:** 
-Tietz NW, Huang WY, Rauh DF, et al. Laboratory tests in the differential diagnosis of hyperamylasemia. Clin Chem 1986 32(2):301-307 +  Tietz NW, Huang WY, Rauh DF, et al. Laboratory tests in the differential diagnosis of hyperamylasemia. Clin Chem 1986 32(2):301-307 
-6 Junge W. Troge B, Klein G, et al. Evaluation of a New Assay for Pancreatic Amylase: Performance Characteristics and Estimation of Reference Intervals Clin Biochem 1989,22 109-114 +  6 Junge W. Troge B, Klein G, et al. Evaluation of a New Assay for Pancreatic Amylase: Performance Characteristics and Estimation of Reference Intervals Clin Biochem 1989,22 109-114 
-Stuttgart/New York Georg Thieme Verlag 1991 354-361. 3 Salt WB II, Schenker S. Amylase its clinical significance: a review ofthe literature [Review] Medicine 1976,55 269-281. +  Stuttgart/New York Georg Thieme Verlag 1991 354-361. 3 Salt WB II, Schenker S. Amylase its clinical significance: a review ofthe literature [Review] Medicine 1976,55 269-281. 
-4 Steinberg WM, Goldstein SS, Davies ND, et al. Diagnostic assays in acute pancreatitis [Review]. Ann Intern Med 1985,102.576-580+  4 Steinberg WM, Goldstein SS, Davies ND, et al. Diagnostic assays in acute pancreatitis [Review]. Ann Intern Med 1985,102.576-580 
 + 
 +|**Printed copy of this document is considered uncontrolled.** It should be compared with controlled electronic copy before use|
  
 +^ Name of Laboratory : Laboratory Services Sir T. Hospital (LSSTH),Bhavnagar ^^^^
 +^Document No.1^**Document Name**: TSH Examination Procedure^**Unique ID**:LSSTH /BIOCHEM/ SOP-5^^
 +^Issue No. : 01^Issue Date :30/04/2024^Page No.^^
 +^Amend No.^ Amend Date ^Prepared by: Section Incharge^Approved & Issued by: HOD,Biochemistry^