Line 3: Line 3:
 =====Header=====  =====Header===== 
 ^Name^Unique ID^Edition^Date of Edition^ ^Name^Unique ID^Edition^Date of Edition^
-|Documentary Procedures for the calibration of equipment |LSSTH/B/Central/DP/6.5.2/9| 1 | 01-03-2023|   +|Documentary Procedures for the calibration of equipment |LSSTH/B/Central/DP/6.5.2/23| 1 | 01-03-2023|   
    
 ^Preparing authority^Approving authority^Review period^Review Date^ ^Preparing authority^Approving authority^Review period^Review Date^
Line 30: Line 30:
 ^1^Centrifuge^(1) Inhouse calibration by Tachometer for speed & time by the Digital stopwatch, Multipoint calibration should be done & evaluated by calculating CV or MU (2) By calibrating agency^Once in year^ ^1^Centrifuge^(1) Inhouse calibration by Tachometer for speed & time by the Digital stopwatch, Multipoint calibration should be done & evaluated by calculating CV or MU (2) By calibrating agency^Once in year^
 ^2^ARCHITECT PLUS C 8000 & ARCHITECT PLUS I1000sr^Calibrated by the manufacturer.During calibration of auto analysers verification of power supply, Light Source, filter, pipettor assembly inclusive of metering pump and syringe, pressure checks wherever applicable, probe alignment and their carry over checks, temperature of temperature-controlled chambers, cuvette calibration (wherever applicable) and system checks should be carried out.^Once a year and after major maintenance.^ ^2^ARCHITECT PLUS C 8000 & ARCHITECT PLUS I1000sr^Calibrated by the manufacturer.During calibration of auto analysers verification of power supply, Light Source, filter, pipettor assembly inclusive of metering pump and syringe, pressure checks wherever applicable, probe alignment and their carry over checks, temperature of temperature-controlled chambers, cuvette calibration (wherever applicable) and system checks should be carried out.^Once a year and after major maintenance.^
-^3^Pipette^(1) Inhouse calibration by gravimetric checks: Repeat at least 10 times volume delivery and weighing. (The laboratory may refer ISO 8655-6:2002 for guidance). (2) By calibrating agency or Biomedical Engineer^Initial manufacturer’s certificate^At the time of purchase and every one year thereafter^+^3^Pipette^(1) Inhouse calibration by gravimetric checks: Repeat at least 10 times volume delivery and weighing. (The laboratory may refer ISO 8655-6:2002 for guidance). (2) By calibrating agency or Biomedical Engineer^At the time of purchase and every one year thereafter^
 ^4^Refrigerator^Refrigerators and deep freezers are verified with calibrated temperature-recording devices for the intended temperature required. An attempt will be made to cover various chambers of the refrigerator and deep freezer. The temperature is monitored with calibrated recording devices at multiple points with the doors shut.^Once a year and after any Corrective actions^ ^4^Refrigerator^Refrigerators and deep freezers are verified with calibrated temperature-recording devices for the intended temperature required. An attempt will be made to cover various chambers of the refrigerator and deep freezer. The temperature is monitored with calibrated recording devices at multiple points with the doors shut.^Once a year and after any Corrective actions^
 ^5^ ^5^