AST estimation from serum NADH (without P-5’-P) Method.

2.Responsibility and Authority:

  • Calibration: Technician
  • Quality Control: Technician
  • Routine operation: Technician
  • Overall Monitoring: Quality Manager

3.Sample Details:

  • Type of Sample: Serum,
  • Type of container and additives: Plain without any additives
  • Patient Preparation: As per Primary Sample Collection Manual sample_collection_manual
  • Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours
  • Handling and transport: As per Primary Sample collection manual
  • Storage: 24 hours at 2-8° C

4.Required Equipment:

  • Centrifuge, Auto-Pipette, Disposable Tips, Disposable sample cups, FullyAuto Chemistry Analyzer

5.Required reagents: R1 L-aspartic acid (103.860 g/L), lactate dehydrogenase (4.000 KU/L), and malate dehydrogenase (2.000 KU/L). R2 α-ketoglutaric acid (6.570 g/L).

6.Reagent Handling

  • Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles
  • Reagent Storage and stability
  • Unopened reagent stable at 2-8°C until expiration date.
  • On board System temperature reagent is stable for 30 days
  • Instability or deterioration should be suspected if there are precipitates, visible signs of leakage or
  • contamination, turbidity, or if calibration or controls do not meet the appropriate criteria.

7.Calibration Procedure:

  • SGOT Calibrators-WATER BLANK
  • Frequency:
  • Reagent lot change
  • QC out of range
  • After service or maintenance
  • Replacement in any parts of Instrument

Procedure:

  • Start the equipment.WDI abbotte fully.docx
  • Calibrators are ready to use.
  • Put calibrator 15-20 minutes at room temperature
  • Prior to use, mix the contents by inverting the vial. Do not shake the vial to prevent foam formation.
  • Take a 150 µl calibrator solution in to separate aliquot.
  • Go to the calibration and give the calibration order.
  • Verify calibration with at least two levels of controls
  • If control results fall outside acceptable ranges, root cause analysis or recalibration may be necessary.

8.Quality control Procedure:

  • Name: Bio Rad Level 1 & 2
  • Frequency: As per Quality Control Procedure

Procedure for Reconstitution of IQC

  • Reconstitution of QC material with 5 ml Distilled water by using calibrated fixed volume pipette.
  • Leave to stand for 30 min in the dark place.
  • Swirl gently several times during the reconstitution period to ensure that the contents are completely dissolved.
  • Prior to use, mix the contents by inverting the vial. Do not shake the vial as the information of foam should be avoided. Ensure that no lyophilized material remains un-reconstituted.
  • Prepare aliquots of 150 µl from the reconstituted QC material.
  • Store these aliquots at -15° C to -20° C.
  • Prior to use, make sure that aliquots should be at room temperature for at least 15 min.

Procedure to run IQC

  1. Press Control order
  2. Select Assay /Panel, to be run.
  3. Select the control/s and its level/s
  4. Give Carrier Number and Position number
  5. Press F3 / Add order
  6. Put respected carrier in RSH rack
  7. Check IQC results, in case outliers call residents.

9.Principle of the procedure used for examinations:

  • AST present in the sample catalyzes the transfer of the amino group from L-aspartate to α-ketoglutarate, forming oxaloacetate and L-glutamate. Oxaloacetate in the presence of nicotinamide adenine dinucleotide (NADH) and malate dehydrogenase (MDH) is reduced to L-malate. In this reaction, NADH is oxidized to NAD+. The reaction is monitored by measuring the rate of decrease in absorbance at 340 nm due to the oxidation of NADH to NAD+.

10.Sample Preparation:

  • Required SampleVolume: 150 µl of the sample
  • Temperature: 37°
  • Take 150-200µl of the sample from Primary tube to the secondary aliquot. Write the sample ID on the aliquot.

Procedure to run Patient sample

  • Press Patient order
  • Select Assay /Panel, to be run.
  • Give Carrier Number and Position number
  • Press F3 / Add order
  • Put respected carrier in RSH rack

11.Performance Characteristics:

  • Linearity: 5 to 4001 U/L
  • The limit of detection (LOD): 4 U/L
  • The limit of quantification(LOQ): 5 U/L
  • Unit:U/L

Normal and critical ranges: Aspartate aminotransferase

male<35U/L
Female<31U/L

12.Laboratory Clinical interpretation: AST is found to be elevated in alcohol-induced liver disease, viral hepatitis, cirrhosis and acute fulminant hepatic failure.

13.Interference and cross reaction: The Following analytes were tested up to the two levels indicated at SGOT concentrations of 30 U/L and 120 U/L, and found not to interfere:

  1. No interference from Bilirubin up to 30 mg/dL
  2. No interference from Hemoglobin up to 62 mg/dL
  3. No interference from Intralipid up to 550 mg/dL
  4. No interference from Sulfapyrdine up to 300 mg/dl
  5. No interference from Sufasalazine up to 300 mg/dl
  6. No interference from Temozolomyde up to20 mg/dl

14.Potential source of variation: Turn around time (TAT): Routine: 6.0 hours Urgent: 2.0 hours

15.Recording of observation:

  • Software backup
  • Machine raw data

16.Storage & Disposal of waste:Follow storage & discard procedure

17.Environmental & Safety control:

  • For in vitro diagnostic use
  • Do not use component before the expiration date
  • Do not mix material from different kit lot numbers.
  • Reagent 2 causes mild skin irritation.

18.References:

  1. Burtis CA,Ashwood ER,editors,Tietz Textbook of clinical chemistry.
  2. World health organization. Biosafety manual,3rd edition.
  3. Clinical and laboratory std institute . Protection of Laboratory workers from occupationally acquired infection; Approved guideline -4th edition.
  4. Burtis CA,Ashwood ER,editors,Tietz Textbook of clinical chemistry
  5. Kaplan LA,Pesce AJ editors. Clinical Chemistry Theory, Analysis, and Correlation, 3rd ed.
Printed copy of this document is considered uncontrolled. It should be compared with controlled electronic copy before use
Name of Laboratory : Laboratory Services Sir T. Hospital (LSSTH),Bhavnagar
Document No.1Document Name: SGOT (AST) Examination ProcedureUnique ID:LSSTH /BIOCHEM/ SOP-5
Issue No. : 01Issue Date :30/04/2024Page No.
Amend No. Amend Date Prepared by: Section InchargeApproved & Issued by: HOD,Biochemistry