This is an old revision of the document!
- Purpose of examination:
- Creatinine estimation from serum or plasma by Kinetic Alkaline Picrate Method.
- Responsibility and Authority:
- Calibration: Technician
- Quality Control: Technician
- Routine operation: Technician
- Overall Monitoring: Quality Manager
- Sample Details:
- Type of Sample: Serum, Plasma
- Type of container and additives: Plain without any additives
- Patient Preparation: As per Primary Sample Collection Manual sample_collection_manual
- Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours
- Handling and transport: As per Primary Sample collection manual
- Storage: 24 hours at 2-8° C
- Required Equipment:
- Centrifuge, Auto-Pipette, Disposable Tips, Disposable sample cups, FullyAuto Chemistry Analyzer
- Required reagents:
- R1. Sodium hydroxide 0.8 mol/L
- R2. Picric acid 24 mmol/L
- Reagent Handling
- Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles
Reagent Storage and stability:
- Unopened reagent stable at 15-30°C until expiration date.
- On board System temperature reagent is stable for 05 days
- Instability or deterioration should be suspected if there are precipitates, visible signs of leakage or contamination, turbidity, or if calibration or controls do not meet the appropriate criteria.
- Calibration Procedure:
- Consolidated Chemistry Calibrators(ConCC)
- Frequency:
- Reagent lot change
- QC out of range
- After service or maintenance
- Replacement in any parts of Instrument
- Procedure:
- Start the equipment.WDI abbotte fully.docx
- Calibrators are ready to use.
- Put calibrator 15-20 minutes at room temperature
- Prior to use, mix the contents by inverting the vial. Do not shake the vial to prevent foam formation.
- Take a 150 µl calibrator solution in to separate aliquot.
- Go to the calibration and give the calibration order.
- Verify calibration with at least two levels of controls
- If control results fall outside acceptable ranges,root cause analysis or recalibration may be necessary.
- Quality control Procedure:
Name: Biorad Level 1 &2
Frequency: As per Quality Control Procedure
- Procedure for Reconstitution of IQC
- Reconstitution of QC material with 5 ml Distilled water by using calibrated fixed volume pipette.
- Leave to stand for 30 min in the dark place.
- Swirl gently several times during the reconstitution period to ensure that the contents are completely dissolved.
- Prior to use, mix the contents by inverting the vial. Do not shake the vial as the information of foam should be avoided. Ensure that no lyophilized material remains un-reconstituted.
- Prepare aliquots of 150 µl from the reconstituted QC material.
- Store these aliquots at -15° C to -20° C.
- Prior to use, make sure that aliquots should be at room temperature for at least 15 min.
- Procedure to run IQC
- Press Control order
- Select Assay /Panel, to be run.
- Select the control/s and its level/s
- Give Carrier Number and Position number
- Press F3 / Add order
- Put respected carrier in RSH rack
- Check IQC results, in case outliers call residents.
- Principle of the procedure used for examinations:
At an alkaline pH, creatinine in the sample reacts with picrate to form a creatinine-picrate complex. The rate of increase in absorbance at 500 nm due to the formation of this complex is directly proportional to the concentration of creatinine in the sample.
- Sample Preparation:
- Required Sample Volume: 150 µl of the sample
- Temperature: 37°
Take 150-200µl of the sample from Primary tube to the secondary aliquot. Write the sample ID on the aliquot.
- Procedure to run Patient sample
- Press Patient order
- Select Assay /Panel, to be run.
- Give Carrier Number and Position number
- Press F3 / Add order
- Put respected carrier in RSH rack
- Performance Characteristics:
- Linearity: 0.20 to 37.00 mg/dL
- The limit of detection (LOD): 0.05 mg/dL
- The limit of quantification(LOQ):0.10 mg/dL
- Unit: mg/dl
- Normal and critical ranges:
1-5 y | 0.04-0.45 | mg/dL |
---|---|---|
6-10 y | 0.22-0.59 | mg/dL |
Adult male | 0.62-1.10 | mg/dL |
Adult Female | 0.45-0.75 | mg/dL |
- Laboratory Clinical interpretation:
- Measurement of serum creatinine is used to diagnose and monitor acute and chronic renal disease, estimate glomerular filtration rate (GFR), or assess the status of renal dialysis patients.
- Interference and cross reaction:
The Following analytes were tested up to the levels indicated at Creatinine concentrations of 0.14mg/dl and 5.03 mg/dl, and found not to interfere:
- No interference from Bilirubin up to 30 mg/dL
- No interference from Hemoglobin up to 1,000 mg/dL
- No interference from Intralipid up to 750 mg/dL
- No interference from Ascorbate up to 1.5 mg/dL
- No interference from Glucose up to 300 mg/dL
- No interference from Protein up to 10.6 g/dL
- Potential source of variation:
- Turn around time (TAT):
- Routine: 6.0 hours
- Urgent: 2.0 hours
- Recording of observation:
- Software backup
- Machine raw data
Storage & Disposal of waste: Follow storage & discard procedure
- Environmental & Safety control:
Reagent R1 contains sodium hydroxide that Causes severe skin burns and eye Damage & May be corrosive to metals
* Precautions:
- Wear protective gloves / protective clothing / eye protection
- Do not breathe mist / vapors / spray
- Wash hands thoroughly after handling
- Keep only in original container * Response * IF SWALLOWED: Rinse mouth. Do NOT induce vomiting * IF IN EYES:Rinse cautiously with water for several minutes. Remove contact lenses, if present and easy to do.Continue rinsing * IF ON SKIN (or hair): Take off immediately all contaminated clothing. Rinse skin with water / shower. For Reagent R2: * Explosive when dry * Picric acid is a flammable solid when wet as a paste (i.e., not less than 10% water), and explosive when dry. * Prevent from forming crystals. * Keep containers tightly sealed. * Do not allow to dry out. * References:**
Thomas L, editor. Clinical Laboratory Diagnostics: Use and Assessment of Clinical Laboratory Results. Frankfurt, Germany: TH- Books Verlagsgesellschaft mbH; 1998:366–374. Jaffe M. Ueber den Niederschlag, welchen Pikrinsaure in normalem Harn erzeugt und uber eine neue Reaction des Kreatinins. Hoppe Seylers Z Physiol Chem 1886;10:391–400. Folin O. Beitrag zur Chemie des Kreatinins und Kreatins Im Harne.Hoppe Seylers Z Physiol Chem 1904;41:223–242. Fabiny DL, Ertingshausen G. Automated reaction-rate method for determination of serum creatinine with the CentrifiChem. Clin Chem 1971;17:696–700. Soldin S, Henderson L, Hill G. The effect of bilirubin and ketones on reaction rate methods for the measurement of creatinine. Clin Biochem 1978:82–86.