SGOT (AST) Examination Procedure
1.Purpose of examination:
AST estimation from serum NADH (without P-5’-P) Method.
2.Responsibility and Authority:
Calibration: Technician
Quality Control: Technician
Routine operation: Technician
Overall Monitoring: Quality Manager
3.Sample Details:
Type of Sample: Serum,
Type of container and additives: Plain without any additives
-
Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours
Handling and transport: As per Primary Sample collection manual
Storage: 24 hours at 2-8° C
4.Required Equipment:
Centrifuge, Auto-Pipette, Disposable Tips, Disposable sample cups, FullyAuto Chemistry Analyzer
5.Required reagents:
R1 L-aspartic acid (103.860 g/L),
lactate dehydrogenase (4.000 KU/L),
and malate dehydrogenase (2.000 KU/L).
R2 α-ketoglutaric acid (6.570 g/L).
6.Reagent Handling
Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles
Reagent Storage and stability
Unopened reagent stable at 2-8°C until expiration date.
On board System temperature reagent is stable for 30 days
Instability or deterioration should be suspected if there are precipitates, visible signs of leakage or
contamination, turbidity, or if calibration or controls do not meet the appropriate criteria.
7.Calibration Procedure:
SGOT Calibrators-WATER BLANK
Frequency:
Reagent lot change
QC out of range
After service or maintenance
Replacement in any parts of Instrument
Procedure:
Start the equipment.WDI abbotte fully.docx
Calibrators are ready to use.
Put calibrator 15-20 minutes at room temperature
Prior to use, mix the contents by inverting the vial. Do not shake the vial to prevent foam formation.
Take a 150 µl calibrator solution in to separate aliquot.
Go to the calibration and give the calibration order.
Verify calibration with at least two levels of controls
If control results fall outside acceptable ranges, root cause analysis or recalibration may be necessary.
8.Quality control Procedure:
Procedure for Reconstitution of IQC
Reconstitution of QC material with 5 ml Distilled water by using calibrated fixed volume pipette.
Leave to stand for 30 min in the dark place.
Swirl gently several times during the reconstitution period to ensure that the contents are completely dissolved.
Prior to use, mix the contents by inverting the vial. Do not shake the vial as the information of foam should be avoided. Ensure that no lyophilized material remains un-reconstituted.
Prepare aliquots of 150 µl from the reconstituted QC material.
Store these aliquots at -15° C to -20° C.
Prior to use, make sure that aliquots should be at room temperature for at least 15 min.
Procedure to run IQC
Press Control order
Select Assay /Panel, to be run.
Select the control/s and its level/s
Give Carrier Number and Position number
Press F3 / Add order
Put respected carrier in RSH rack
Check IQC results, in case outliers call residents.
9.Principle of the procedure used for examinations:
AST present in the sample catalyzes the transfer of the amino group from L-aspartate to α-ketoglutarate, forming oxaloacetate and L-glutamate. Oxaloacetate in the presence of nicotinamide adenine dinucleotide (NADH) and malate dehydrogenase (MDH) is reduced to L-malate. In this reaction, NADH is oxidized to NAD+. The reaction is monitored by measuring the rate of decrease in absorbance at 340 nm due to the oxidation of NADH to NAD+.
10.Sample Preparation:
Procedure to run Patient sample
Press Patient order
Select Assay /Panel, to be run.
Give Carrier Number and Position number
Press F3 / Add order
Put respected carrier in RSH rack
11.Performance Characteristics:
Normal and critical ranges:
Aspartate aminotransferase
12.Laboratory Clinical interpretation:
AST is found to be elevated in alcohol-induced liver disease, viral hepatitis, cirrhosis and acute
fulminant hepatic failure.
13.Interference and cross reaction:
The Following analytes were tested up to the two levels indicated at SGOT concentrations of 30 U/L and 120 U/L, and found not to interfere:
No interference from Bilirubin up to 30 mg/dL
No interference from Hemoglobin up to 62 mg/dL
No interference from Intralipid up to 550 mg/dL
No interference from Sulfapyrdine up to 300 mg/dl
No interference from Sufasalazine up to 300 mg/dl
No interference from Temozolomyde up to20 mg/dl
14.Potential source of variation:
Turn around time (TAT):
Routine: 6.0 hours
Urgent: 2.0 hours
15.Recording of observation:
Software backup
Machine raw data
16.Storage & Disposal of waste:Follow storage & discard procedure
17.Environmental & Safety control:
For in vitro diagnostic use
Do not use component before the expiration date
Do not mix material from different kit lot numbers.
Reagent 2 causes mild skin irritation.
18.References:
Burtis CA,Ashwood ER,editors,Tietz Textbook of clinical chemistry.
World health organization. Biosafety manual,3rd edition.
Clinical and laboratory std institute . Protection of Laboratory workers from occupationally acquired infection; Approved guideline -4th edition.
Burtis CA,Ashwood ER,editors,Tietz Textbook of clinical chemistry
Kaplan LA,Pesce AJ editors. Clinical Chemistry Theory, Analysis, and Correlation, 3rd ed.
Printed copy of this document is considered uncontrolled. It should be compared with controlled electronic copy before use |
Name of Laboratory : Laboratory Services Sir T. Hospital (LSSTH),Bhavnagar |
Document No.1 | Document Name: SGOT (AST) Examination Procedure | Unique ID:LSSTH /BIOCHEM/ SOP-5 |
Issue No. : 01 | Issue Date :30/04/2024 | Page No. |
Amend No. | Amend Date | Prepared by: Section Incharge | Approved & Issued by: HOD,Biochemistry |