Chloride Examination Procedure
1.Purpose of Examination: To lay down standard operating procedure for serum Chloride by ISE Method in Biochemistry section of LSSTH, Bhavnagar.
2.Principle: Ion Selective Electrode.
3.Performance: Linearity – 70-152 mEq/L Precision - It is determined from 2 runs per day with two replicates per run for 20 days on two AFT 300 Electrolyte Analyzer.
4.Primary Sample: Plasma or serum sample_collection_manual
5.Patient Preparation: Verbal consent of patient should be taken before collecting blood sample.
6.Type of Container: Plain Vaccute with no additive
7.Required Equipment: Centrifuge, Electrolyte auto analyzer
8.Safety Precaution: Follow the universal work precautions.
9.Calibration Procedure: Instrument performs auto calibration every four hours Manual calibration is done on following circumstances:
10.Procedural Steps: Centrifuge the blood sample ,Open the sample door of the instrument, than hold the Vacutte under the probe to be aspirated
11.Quality Control Procedure: The Laboratory runs Internal Quality control level 1, level 2 & level 3 twice a day. Monthly L-J charts are reviewed & for any QC outlier root cause analysis are done & corrective actions are taken according to error. Participated in RIQAS programme. RIQAS report is reviewed and if any result go outlier, than root cause analysis is done & corrective actions is taken according to error.
12.Interferences & Cross-Reactions: Salicylates in high level, Lipemia and Hemolysis
13.Principle of Procedure for Calculating Results: Membrane potential is caused by the permeability of certain types of membranes to selected anions or cations. The potential produced at the membrane –solution interface is proportional to the logarithm of the ionic activity or concentration of Chloride Ion.
14.Biological Reference Interval: 96-106 mEq/L
15.Reportable interval of patient examination results: 80-120 mEq/L
16.Instruction for determining Quantitative results when a results is not within the measurement Rerun the sample and correlate clinically
17.Stability of sample Serum or plasma-At Room temperature18°–28°C (64°–82°F) stability ≤ 24 hours CSF- At Refrigerated 2°–8°C (36°–46°F) stability 3 days
18.Storage of sample 2-8° C for 1 days (Retention period for re-examination and/or additional tests is 24 hrs. at (2-8° C).
19.Reagent Storage and stability Unopened reagents is stable until expiration date when store at 2-8°c.Reagent stability is 30 days if reagent is uncapped and onboard.
20.Turn around time (TAT) 6.0 hours after collection, in case of emergency 2.0 hours.
21.Critical Value: 80-120 mEq/L
22.Laboratory Interpretation:
Dehydration, Cushing’s Syndrome, Hyperventilation and eclampsia
Diarrhoea, Vomitting, Ulcerative colitis, Pyloric Obstruction, Severe burns, Diabetic acidosis, Addison’s disease
23.Potential source of variation:
24.Reagent Handling Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to sit at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles
25.REFERENCE:
Printed copy of this document is considered uncontrolled. It should be compared with controlled electronic copy before use |
Name of Laboratory : Laboratory Services Sir T. Hospital (LSSTH),Bhavnagar | |||
---|---|---|---|
Document No.1 | Document Name: Chloride Examination Procedure | Unique ID:LSSTH /BIOCHEM/ SOP-5 | |
Issue No. : 01 | Issue Date :30/04/2024 | Page No. | |
Amend No. | Amend Date | Prepared by: Section Incharge | Approved & Issued by: HOD,Biochemistry |