Total Bilirubin Examination Procedure
1.Purpose of examination:
2.Responsibility and Authority:
Calibration: Technician
Quality Control: Technician
Routine operation: Technician
Overall Monitoring: Quality Manager
3.Sample Details:
Type of Sample: Serum, Plasma
Type of container and additives: Plain without any additives
-
Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours
Handling and transport: As per Primary Sample collection manual
Storage: 24 hours at 2-8° C
4.Required Equipment:
Centrifuge, Auto-Pipette, Disposable Tips, Disposable sample cups, FullyAuto Chemistry Analyzer
5.Required reagents:
Hydrochloric acid 9.33 g/L
Hydrochloric acid 5.563 g/L
Sodium nitrite 0.345 g/L
Surfactant 1.96%
6.Reagent Handling
Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles
7.Reagent Storage and stability
Unopened reagent stable at 2-8°C until expiration date.
On board System temperature reagent is stable for 21 days
Instability or deterioration should be suspected if there are precipitates, visible signs of leakage or
contamination, turbidity, or if calibration or controls do not meet the appropriate criteria.
8.Calibration Procedure:
Billi Calibrators- Level 1 & 2
Frequency:
Reagent lot change
QC out of range
After service or maintenance
Replacement in any parts of Instrument
9.Procedure:
Start the equipment.WDI abbotte fully.docx
Calibrators are ready to use.
Put calibrator 15-20 minutes at room temperature
Prior to use, mix the contents by inverting the vial. Do not shake the vial to prevent foam formation.
Take a 150 µl both level of calibrator solution in to separate aliquot.
Go to the calibration and give the calibration order.
Verify calibration with at least two levels of controls
If control results fall outside acceptable ranges,root cause analysis or recalibration may be necessary.
10.Quality control Procedure:
Procedure for Reconstitution of IQC
Reconstitution of QC material with 5 ml Distilled water by using calibrated fixed volume pipette.
Leave to stand for 30 min in the dark place.
Swirl gently several times during the reconstitution period to ensure that the contents are completely dissolved.
Prior to use, mix the contents by inverting the vial. Do not shake the vial as the information of foam should be avoided. Ensure that no lyophilized material remains un-reconstituted.
Prepare aliquots of 150 µl from the reconstituted QC material.
Store these aliquots at -15° C to -20° C.
Prior to use, make sure that aliquots should be at room temperature for at least 15 min.
Procedure to run IQC
Press Control order
Select Assay /Panel, to be run.
Select the control/s and its level/s
Give Carrier Number and Position number
Press F3 / Add order
Put respected carrier in RSH rack
Check IQC results, in case outliers call residents.
11.Principle of the procedure used for examinations:
Total (conjugated and unconjugated) bilirubin couples with a diazo
reagent in the presence of a surfactant to form azobilirubin. The diazo
reaction is accelerated by the addition of surfactant as a solubilizing
agent. The increase in absorbance at 548 nm due to azobilirubin is
directly proportional to the total bilirubin concentration.
12.Sample Preparation:
Procedure to run Patient sample
Press Patient order
Select Assay /Panel, to be run.
Give Carrier Number and Position number
Press F3 / Add order
Put respected carrier in RSH rack
13.Performance Characteristics:
Linearity: 0.1 to 25.0 mg/dL
If values exceed this linearity limit 25.00 mg/dl, dilute the sample by Manual Dilution Procedure, or the Automatic Dilution Protocol provided in the assay parameters.
Automated Dilution Protocol:
When using the Automated Dilution Protocol, the system performs a dilution of the specimen and automatically corrects the concentration by multiplying the result by the appropriate dilution factor.
Manual Dilution Procedure:
Dilute the specimen with saline (0.85% to 0.90% NaCl). Enter the dilution factor in the Patient or Control order screen
Normal and critical ranges:
0-1 d | 2.0-6.0 | mg/dL |
1-2 d | 6.0-10.0 | mg/dL |
3-5 d | 4.0-8.0 | mg/dL |
1 week -18 years | 0-2.06 | mg/dL |
Adult | 0.2-1.5 | mg/dL |
14.Laboratory Clinical interpretation:
Unconjugated hyper-bilirubinemia is seen in newborns, in increased red cell destruction (hemolytic anemia, extension hematoma), is ineffective erythropoiesis some rare genetic disease (Gilbert's syndrome, Crigler- Najjar syndrome)
Conjugated hyper-bilirubinemiais associated to a decreased excretion of bile due to liver disease (hepatitis or cirrhosis) or to intrahepatic or extrahepatic cholestasis
15.Interference and cross reaction:
The Following analytes were tested up to the levels indicated at Direct Bilirubin concentrations of 0.14mg/dl and 5.03 mg/dl, and found not to interfere:
No interference from Hemoglobin up to 1,000 mg/dL
No interference from Intralipid up to 750 mg/dL
No interference from Indican up to 0.175 mmol/L
No interference from Indocynine green up to 18.8 mg/dL
16.Potential source of variation:
Routine: 6.0 hours
Urgent: 2.0 hours
17.Recording of observation:
Software backup
Machine raw data
18.Storage & Disposal of waste: Follow storage & discard procedure
19.Environmental & Safety control:
Reagent R1 contains hydrochloric acid and sodium borohydride. Reagent R2 contains hydrochloric acid.
that Causes severe skin burns and eye damage.
May be corrosive to metal
20.Precautions:
Keep only in original container.
Do not breathe mist/vapors/spray.
Wash hands thoroughly after handling.
Wear protective gloves/protective
21.References:
Jacobs DS, DeMott WR, Grady HJ, et al. Laboratory Test Handbook, 4th ed. Hudson, OH: Lexi‑Comp; 1996:86.
Dennery PA, Seidman DS, Stevenson DK. Drug therapy: neonatal hyperbilirubinemia. N Engl J Med 2001;344(8):581–90.
Malloy HT, Evelyn KA. The determination of bilirubin with the photoelectric colorimeter. J Biol Chem 1973;119:481–90.
Burtis CA, Ashwood ER, editors. Tietz Textbook of Clinical Chemistry, 3rd ed. Philadelphia, PA: WB Saunders; 1999:1136–7.
Walters MI, Gerarde HW. An ultramicromethod for the determination of conjugated and total bilirubin in serum or plasma. Microchemical J 1970;15:231–43.
ARCHITECT TOTAL BILI 6L45 307147/R04
NABL 112 effective from 01/06/2019
Printed copy of this document is considered uncontrolled. It should be compared with controlled electronic copy before use |
Name of Laboratory : Laboratory Services Sir T. Hospital (LSSTH),Bhavnagar |
Document No.1 | Document Name: Total Bilirubin Examination Procedure | Unique ID:LSSTH /BIOCHEM/ SOP-5 |
Issue No. : 01 | Issue Date :30/04/2024 | Page No. |
Amend No. | Amend Date | Prepared by: Section Incharge | Approved & Issued by: HOD,Biochemistry |