|[[clinical_biochemistry_section|Home]]|[[clinical_biochemistry|]]|[[examination_procedures|]]| Ferritin Examination Procedure **1.Purpose of examination**: Ferritin estimation from serum or plasma by CMIA technology (Chemiflex) Method. **2.Responsibility and Authority:** * Calibration: Technician * Quality Control: Technician * Routine operation: Technician * Overall Monitoring: Quality Manager **3.Sample Details:** - Type of Sample: Serum, Plasma - Type of container and additives: Plain without any additives - Patient Preparation: As per Primary Sample Collection Manual [[sample_collection_manual|]] - Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours - Handling and transport: As per Primary Sample collection manual - Storage: 24 hours at 2-8° C **4.Required Equipment:** Centrifuge, Auto-Pipette, Disposable Tips, Disposable sample cups, FullyAuto Chemistry Analyzer **5.Required reagents:** Microparticles: Anti-Ferritin (mouse, monoclonal) coated Microparticles in TRIS buffer with protein (mouse and bovine) stabilizers. Minimum Concentration: 0.125% solids. Conjugate : Anti-Ferritin (rabbit, polyclonal) acridinium labeled Conjugate in MES buffer with protein (bovine) stabilizers. Minimum concentration: 75 ng/mL. Multi Assay manual dilution : phosphate buffered saline solution. Pre trigger solution: 1.32% (w/v) hydrogen peroxide. Trigger solution : 0.35 N sodium hydroxide. Wash-buffer : phosphate buffered saline solution. **6.Reagent Handling** Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles **7.Reagent Storage and stability** * Unopened reagent stable at 15-30°C until expiration date. * On board System temperature reagent is stable for 30 days * Instability or deterioration should be suspected if there are precipitates, visible signs of leakage or * contamination, turbidity, or if calibration or controls do not meet the appropriate criteria. **8.Calibration Procedure:** * Ferritin Calibrators * Frequency: - Reagent lot change - QC out of range - After service or maintenance - Replacement in any parts of Instrument **Procedure:** - Start the equipment.WDI abbotte fully.docx - Calibrators are ready to use. - Put calibrator 15-20 minutes at room temperature - Prior to use, mix the contents by inverting the vial. Do not shake the vial to prevent foam formation. - Take a 150 µl calibrator solution in to separate aliquots. - Go to the calibration and give the calibration order. - Verify calibration with at least two levels of controls - If control results fall outside acceptable ranges,root cause analysis or recalibration may be necessary. **9.Quality control Procedure:** Name: Bio Rad Level 1 , 2 & 3 Frequency: As per Quality Control Procedure * Procedure for Reconstitution of IQC * Reconstitution of QC material with 5 ml Distilled water by using calibrated fixed volume pipette. * Leave to stand for 30 min in the dark place. * Swirl gently several times during the reconstitution period to ensure that the contents are completely dissolved. * Prior to use, mix the contents by inverting the vial. Do not shake the vial as the information of foam should be avoided. Ensure that no lyophilized material remains un-reconstituted. * Prepare aliquots of 150 µl from the reconstituted QC material. * Store these aliquots at -15° C to -20° C. * Prior to use, make sure that aliquots should be at room temperature for at least 15 min. Procedure to run IQC - Press Control order - Select Assay /Panel, to be run. - Select the control/s and its level/s - Give Carrier Number and Position number - Press F3 / Add order - Put respected carrier in RSH rack - Check IQC results, in case outliers call residents. **10.Principle of the procedure used for examinations:** - Sample, and anti-ferritin coated paramagnetic microparticles are combined. The ferritin present in the sample binds to the anti- ferritin coated microparticles. - After washing, anti-ferritin acridinium-labeled conjugate is added to create a reaction mixture. - Following another wash cycle, Pre-Trigger and Trigger Solutions are added to the reaction mixture. - The resulting chemiluminescent reaction is measured as relative light units (RLUs). There is a direct relationship between the amount of ferritin in the sample. **11.Sample Preparation:** Required SampleVolume: 150 µl of the sample Temperature: 37° Take 150-200µl of the sample from Primary tube to the secondary aliquot. Write the sample ID on the aliquot. **12.Procedure to run Patient sample** * Press Patient order * Select Assay /Panel, to be run. * Give Carrier Number and Position number * Press F3 / Add order * Put respected carrier in RSH rack **13.Performance Characteristics:** * Linearity: ng/mL * The limit of detection (LOD): ng/mL * The limit of quantification(LOQ): ng/mL * Unit: ng/mL ** Normal and critical ranges:** ^1 mo^200-600^ng/mL^ ^0-1 Y^50-200^ng/mL^ ^1-15 y^7-140^ng/mL^ ^Male^20-250^ng/mL^ ^Female^10-120^ng/mL^ **14.Laboratory Clinical interpretation:** Specifically, the combined use of serum ferritin levels and mean corpuscular volume (MCV) has made differentiation between iron deficiency, beta-thalassemia trait and normal subjects possible at a very high level of accuracy. **15.Interference and cross reaction:** ARCHITECT Ferritin assay demonstrated ≤ 10% mean interference at the levels indicated below. * Hemoglobin 200 mg/dL * Bilirubin 20 mg/dL * Triglycerides 3000 mg/dL * Protein 2 g/dL and 12 g/dL **16.Potential source of variation:** Turn around time (TAT): Routine: 6.0 hours Urgent: 2.0 hours **17.Recording of observation:** * Software backup * Machine raw data **18.Storage & Disposal of waste:** Follow storage & discard procedure **19.Environmental & Safety control**: Contact with acids liberates very toxic gas. Dispose of contents / container in accordance with local regulations. **20.References:** - Krause JR, Stolc V. Serum Ferritin and Bone Marrow Iron Stores - Correlation with Absence of Iron in Biopsy Specimens. Am J ClinPathol 1979;72:817-820. - Skikne BS, Cook JD. Serum Ferritin in the Evaluation of Iron Status.Lab Management 1981;19:31-35. - Addison GM, Beamish MR, Hales CN, et al. An ImmunoradiometricAssay for Ferritin in the Serum of Normal Subjects and Patients withIron Deficiency and Iron Overload. J Clin Pathol 1972;25:326-329. - Jacobs A, Miller F, Worwood M, et al. Ferritin in the Serum of NormalSubjects and Patients with Iron Deficiency and Iron Overload. Br MedJ 1972;4:206-208. - Lipschitz DA, Cook JD, Finch CA. A Clinical Evaluation of SerumFerritin as an Index of Iron Stores. ^ Name of Laboratory : Laboratory Services Sir T. Hospital (LSSTH),Bhavnagar ^^^^ ^Document No.1^**Document Name**: Ferritin Examination Procedure^**Unique ID**:LSSTH /BIOCHEM/ SOP-5^^ ^Issue No. : 01^Issue Date :30/04/2024^Page No.^^ ^Amend No.^ Amend Date ^Prepared by: Section Incharge^Approved & Issued by: HOD,Biochemistry^