Differences
This shows you the differences between two versions of the page.
Next revision | Previous revision | ||
total_cholesterol [2025/01/27 17:31] – created admin | total_cholesterol [2025/01/28 05:01] (current) – admin | ||
---|---|---|---|
Line 1: | Line 1: | ||
+ | |[[clinical_biochemistry_section|Home]]|[[clinical_biochemistry|]]|[[examination_procedures|]]| | ||
+ | Total Cholesterol Examination Procedure | ||
+ | **1.Purpose of Examination: | ||
+ | To lay down standard operating procedure for serum Total Cholesterol estimation by CHOD-PAP Method in Biochemistry section of LSSTH, Bhavnagar. | ||
- | Purpose | + | **2.Principle |
- | ●Triglceride estimation from serum or plasma | + | The Cholesterol2 assay is an automated clinical chemistry assay. Cholesterol esters are enzymatically hydrolyzed |
+ | (quinoneimine dye) which is quantitated at 604 nm. | ||
- | Responsibility and Authority: | + | **3.Performance:** |
- | ● Calibration: Technician | + | Linearity: upto 705mg/dl |
- | ● Quality Control: Technician | + | Limit of Detection (LOD):Serum : 5.0 mg/dl |
- | ● Routine operation: Technician | + | Limit of Quantitation (LOQ),:Serum :6.2 mg/dl |
- | ● Overall Monitoring: Quality Manager | + | The LOQ is the analyte concentration at which the CV = 20%. |
- | | + | **4.Type of Primary |
- | ● Type of Sample: Serum, | + | Serum or Plasma |
- | ● Type of container and additives: Plain without any additives | + | |
- | ● Patient Preparation: | + | |
- | ● Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours | + | |
- | ● Handling and transport: As per Primary Sample collection manual | + | |
- | ● Storage: 24 hours at 2-8° C | + | |
- | Required Equipment: | + | **5.Patient Preparation:** |
- | ●Centrifuge, | + | Instruction should be given to the patient |
+ | Verbal consent of patient should be taken before collecting blood sample. | ||
+ | Patient should relax for about 5 minutes before the venepuncture.. | ||
- | Required reagents: | + | **6.Type of Container and additives |
- | R1 84ml Active ingredients: | + | Plain Vaccutainer with no additives |
- | | + | |
- | | + | |
- | GPO (microbial) (> 6000 U/L), Lipoprotein lipase (microbial) (> 3000 U/L). | + | |
- | | + | **7.Required Equipment |
- | ● Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles | + | Centrifuge, Auto-Pipette, Disposable Tips, Disposable sample cups, Fully Auto Chemistry Analyzer |
- | Reagent Storage | + | |
- | Unopened reagent stable at 2-8°C until expiration date. | + | |
- | On board System temperature reagent is stable for 30 days | + | |
- | Instability or deterioration should be suspected if there are precipitates, visible signs of leakage or | + | |
- | contamination, turbidity, or if calibration or controls do not meet the appropriate criteria. | + | |
+ | **8.Enviromental & Safety control:** | ||
+ | Reagent R1 contains sodium azide. contact with acids liberates very toxic gas | ||
+ | **9.Precautions: | ||
+ | Wear protective gloves / protective clothing / eye protection | ||
+ | Do not breathe mist / vapors / spray | ||
+ | Wash hands thoroughly after handling | ||
+ | Keep only in original container | ||
- | Calibration Procedure: | + | **10.Calibration Procedure:** |
- | ● Consolidated Chemistry Calibrator | + | Give Calibration with Consolidated Chemistry |
+ | **Manual calibration is done on following circumstances: | ||
+ | * After instrument repair | ||
+ | * After new reagent lot | ||
+ | * QC data not satisfactory - QC outlier | ||
+ | Note: Verify calibration with at least two levels of controls according to the established quality control requirements for your laboratory. If control results fall outside acceptable ranges, recalibration may be necessary | ||
- | ● Frequency: | + | **11.Procedural Steps:** |
- | Reagent lot change | + | * Verify that all necessary reagents present |
- | QC out of range | + | * Take100µl Sample. |
- | After service or maintenance | + | |
- | Replacement in any parts of Instrument | + | * Press run. |
- | Procedure: | + | |
- | Start the equipment.WDI abbotte fully.docx | + | |
- | Calibrators are ready to use. | + | |
- | Put calibrator 15-20 minutes at room temperature | + | |
- | Prior to use, mix the contents by inverting the vial. Do not shake the vial to prevent foam formation. | + | |
- | Take a 150 µl calibrator solution in separate aliquots. | + | |
- | Go to the calibration and give the calibration order. | + | |
- | Verify calibration with at least two levels of controls. | + | |
- | If control results fall outside acceptable ranges,root cause analysis or recalibration may be necessary. | + | |
- | Quality | + | **12.Quality |
- | Name: Bio Rad Level 1 & 2 | + | The Laboratory runs Lypocheck Assayed Chemistry Control level 1 & level 2 Once a day. Monthly L-J charts are reviewed & for any QC outlier root cause analysis are done & corrective actions are taken according |
- | Frequency: As per Quality Control Procedure | + | The Laboratory Participates in EQAS programme EQAS report is reviewed and if any result go outlier, then root cause analysis is done & corrective actions |
- | ●Procedure | + | |
- | Reconstitution of QC material with 5 ml Distilled water by using calibrated fixed volume pipette. | + | |
- | Leave to stand for 30 min in the dark place. | + | |
- | Swirl gently several times during the reconstitution period to ensure that the contents | + | |
- | Prior to use, mix the contents by inverting the vial. Do not shake the vial as the information of foam should be avoided. Ensure that no lyophilized material remains un-reconstituted. | + | |
- | Prepare aliquots of 150 µl from the reconstituted QC material. | + | |
- | Store these aliquots at -15° C to -20° C. | + | |
- | Prior to use, make sure that aliquots should be at room temperature for at least 15 min. | + | |
+ | **13.Interferences & Cross-Reactions: | ||
+ | InterferingSubstance | ||
+ | * No interference from Bilirubin up to7.5 mg/dL | ||
+ | * No interference from Hemoglobin up to750 mg/dL | ||
+ | * No interference from Intralipid up to 1000 mg/dl | ||
+ | * No interference from Ascorbate up to 1.5 mg/dl | ||
+ | **14.Principle of Procedure for Calculating Results including where relevant, the measurement uncertainty of measured quantity values:** | ||
+ | Instrument measures the absorbance of a calibrator of known concentration. This point determines the slope of a straight line that relates absorbance to concentration. Instrument calculates a constant which is the inverse of the slope of the straight line. | ||
+ | Sample Concentration: | ||
+ | Cs = Sample Concentration | ||
+ | K = Constant | ||
+ | As = Sample Absorbance | ||
+ | Measurement uncertainty: | ||
+ | Highest CV% of last 6 months | ||
+ | Refer Annexure I | ||
+ | **15.Biological Reference Interval:** | ||
+ | Adult : < | ||
+ | **16.Reportable interval of patient examination results:** | ||
+ | Up to 500mg/dl | ||
+ | * Instruction for determining Quantitative results when a results is not within the measurement | ||
+ | * Serum and plasma specimens with Cholesterol values exceeding the 705mg/dl, dilution need by following the Manual Dilution Procedure, or the Automatic Dilution Protocol provided in the assay parameters | ||
+ | * Automated Dilution Protocol | ||
+ | * When using the Automated Dilution Protocol, the system performs a dilution of the specimen and automatically corrects the concentration by multiplying the result by the appropriate dilution factor. | ||
- | ●Procedure | + | **17.Manual Dilution |
- | Press Control order | + | * Dilute the specimen with saline (0.85% |
- | Select Assay / | + | * Enter the dilution factor in the Patient or Control |
- | Select | + | |
- | Give Carrier Number and Position number | + | |
- | Press F3 / Add order | + | |
- | Put respected carrier in RSH rack | + | |
- | Check IQC results, in case outliers call residents. | + | |
- | Principle | + | **18.Stability |
- | Triglycerides are enzymatically hydrolyzed by lipase to free fatty acids and glycerol. The glycerol is phosphorylated by adenosine triphosphate (ATP) with glycerol kinase (GK) to produce glycerol-3-phosphate and adenosine diphosphate | + | Serum or plasma-At Room temperature18°–28°C |
- | | + | **19.Storage |
- | ● Required SampleVolume: | + | 2-8° C for 1 days (Retention period for re-examination and/or additional tests is 24 hrs.) |
- | ● Temperature: | + | |
- | Take 150-200µl of the sample from Primary tube to the secondary aliquot. Write the sample ID on the aliquot. | + | **20.Reagent Storage and stability** |
- | ●Procedure to run Patient sample | + | Reagent is stored at 2-8 °c.Reagent stability is 30 days if the reagent is uncapped |
- | Press Patient order | + | |
- | Select Assay /Panel, to be run. | + | |
- | Give Carrier Number | + | |
- | Press F3 / Add order | + | |
- | Put respected carrier in RSH rack | + | |
- | Performance Characteristics: | + | **21.Turn around time (TAT)** |
- | ● Linearity: up to 1420 mg/dL | + | 6.0 hours after collection, |
- | ● The limit of detection | + | |
- | ● The limit of quantification(LOQ): | + | |
- | ● Unit: mg/dL | + | |
- | | + | **22.Critical Value:**NA |
- | Triglyceride | + | **23.Laboratory Interpretation: |
- | Desirable | + | The common causes of high cholesterol level are as follows : |
- | Borderline High | + | |
- | High | + | |
- | <200 | + | |
- | 200-399 | + | |
- | 400 and above | + | |
+ | **24.Potential source of variation: | ||
+ | - Expiry of Kit | ||
+ | - Instrumental Error | ||
+ | - Presence of interfering source | ||
+ | - Instrumental Error | ||
+ | - Reagent contamination | ||
+ | - Storage condition not proper | ||
- | | + | **25.Reagent Handling** |
- | The common causes of hyperlipidemia is nephrosis,DM, | + | Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to sit at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles |
- | Interference | + | **26.REFERENCE: |
- | InterferingSubstance | + | - US department of labor, |
- | No interference from Bilirubin up to7.5 mg/dL | + | - US department of health and human services. Biosafety in MIcrobiological and biomedical laboratories. |
- | No interference from Hemoglobin up to750 mg/dL | + | - World health organization. Biosafety manual,3rd edition. |
- | No interference from Ascorbate up to 1.5 mg/dl | + | - World health organization. Biosafety manual,3rd edition. |
- | No interference | + | - Clinical and laboratory std institute . Protection of Laboratory workers |
- | No interference from Dipyrone up to100mg/dl. | + | - Burtis CA,Ashwood ER, |
- | No interference | + | - ARCHITECT |
+ | - NABL 112 effective | ||
- | Potential source | + | |**Printed copy of this document is considered uncontrolled.** It should be compared |
- | Turn around time (TAT): | + | |
- | Routine: 6.0 hours | + | |
- | Urgent: 2.0 hours | + | |
- | + | ||
- | | + | |
- | Software backup | + | |
- | Machine raw data | + | |
- | Storage & Disposal of waste: Follow storage & discard procedure | + | |
- | Environmental & Safety control: | + | |
- | Reagent R1 contains sodium azide.contact | + | |
- | Precautions: | + | |
- | Wear protective gloves / protective clothing / eye protection | + | |
- | Do not breathe mist / vapors / spray | + | |
- | Wash hands thoroughly after handling | + | |
- | Keep only in original container | + | |
- | . | + | |
- | References: | + | |
- | US department | + | ^ Name of Laboratory : Laboratory Services Sir T. Hospital (LSSTH),Bhavnagar ^^^^ |
- | US department of health and human services. Biosafety in MIcrobiological and biomedical laboratories. | + | ^Document No.1^**Document Name**: Total Cholesterol Examination Procedure^**Unique ID**:LSSTH /BIOCHEM/ SOP-5^^ |
- | World health organization. Biosafety manual,3rd edition. | + | ^Issue No. : 01^Issue Date : |
- | World health organization. Biosafety manual,3rd edition. | + | ^Amend No.^ Amend Date ^Prepared by: Section Incharge^Approved |
- | Clinical and laboratory std institute . Protection of Laboratory workers from oocupationally acquired infection; | + | |
- | Burtis CA,Ashwood ER, | + | |