[[quality_manual|QM home]] | [[quality_manual|QM home]] | ||
^Name^Unique ID^Edition^Date of Edition^ | ^Name^Unique ID^Edition^Date of Edition^ | ||
- | |Documentary procedure for protein electrophoresis|LSSTH/ | + | |Documentary procedure for protein electrophoresis|LSSTH/ |
^Preparing authority^Approving authority^Review period^ | ^Preparing authority^Approving authority^Review period^ | ||
Line 47: | Line 47: | ||
- After 10 minutes removes siliconized glass slide by sliding movement and take care so that gel does not broken down. | - After 10 minutes removes siliconized glass slide by sliding movement and take care so that gel does not broken down. | ||
- Gel is ready for sample application. | - Gel is ready for sample application. | ||
+ | - Immediately apply the sample to prevent excess drying. | ||
+ | **Sample Applicator Preparation** | ||
+ | Prepare comb from the stainless steel as shown in image. | ||
+ | Attach the magnetic strip to the ruler. | ||
+ | Stick the ruler with the platform by feviquick. | ||
+ | Stick the Applicator to magnetic strip. Applicator can be move up and down. | ||
| | ||
**Sample application** | **Sample application** | ||
+ | - Put Rough non sliconized glass slide below the sample applicator. | ||
+ | - Apply 4 micro liter sample over the sample applicator at the edge of applicator as shown in figure. so that excess sample will be removed on glass slide. | ||
+ | - Lift up the sample applicator and wipe the back side of applicator so that backward running of sample can be prevented. | ||
+ | - Replace the rough glass slide with glass slide having prepared gel. | ||
+ | - Apply the sample near the one end of gel with the applicator and wait for 2 minutes to allow the samples to get completely absorbed within the gel. | ||
+ | - Remove the applicator. | ||
+ | | ||
+ | **Running the samples in electrophoresis tank** | ||
+ | |||
+ | - Clean the electrophoresis tank with DI water and dry it with tissue paper before use. | ||
+ | - Put the electrophoresis tank on a table having even & leveled surface. | ||
+ | - Fill the cold buffer in the two compartment. Put the gel slide in electrophoresis chamber in such a way that sample applied should remain on cathodic side. | ||
+ | - Connect the gel with buffer through strip of filter paper wick & cover the chamber with its lid. | ||
+ | - Connect the electrode on the side of slide containing sample well to the cathode (-) & other electrode to the anode of the power supply. | ||
+ | - Set voltage at 300 volt and & run for 10 minutes. | ||
+ | - After 10 minutes, | ||
+ | |||
+ | **Fixation** | ||
+ | |||
+ | Put the slide in to methanol for 10 minutes for fixation so that proteins get denatured and remain at same position even after staining and destaining. | ||
+ | |||
+ | **Drying of gel** | ||
+ | |||
+ | Put the fixated glass slide in to hot air Owen at 50-60° C for 20 minutes. | ||
+ | |||
+ | **Staining** | ||
+ | |||
+ | Place the gel slide in to Amido black staining solution for 2 minutes. | ||
+ | |||
+ | **Destaining** | ||
+ | |||
+ | Remove the excess stain by putting the gel in to destaining solution for 15 minutes so that background becomes clear. | ||
+ | |||
+ | **Interpretation of the result** | ||
+ | |||
+ | |||
- | Apply 4 micro liter sample over the sample applicator at the edge of applicator as shown in figure. | ||