Differences
This shows you the differences between two versions of the page.
Next revision | Previous revision | ||
amylase [2025/01/28 05:54] – created admin | amylase [2025/01/28 05:59] (current) – admin | ||
---|---|---|---|
Line 5: | Line 5: | ||
**1.Purpose of examination: | **1.Purpose of examination: | ||
- | ● Amylase | + | Amylase |
- | 2.Responsibility and Authority: | + | **2.Responsibility and Authority:** |
- | ● Calibration: | + | |
- | ● Quality Control: Technician | + | |
- | ● Routine operation: Technician | + | |
- | ● Overall Monitoring: Quality Manager | + | |
- | 3.Sample Details: | + | **3.Sample Details:** |
- | ● Type of Sample: Serum, | + | |
- | ● Type of container and additives: Plain without any additives | + | |
- | ● Patient Preparation: | + | |
- | ● Stability: At Room temperature 18°–28°C (64°–82°F) stability ≤ 24 hours | + | |
- | ● Handling and transport: As per Primary Sample collection manual | + | |
- | ● Storage: 24 hours at 2-8° C | + | |
- | Required Equipment: | + | **4.Required Equipment:** |
- | ●Centrifuge, Auto-Pipette, | + | Centrifuge, Auto-Pipette, |
- | 4.Required reagents: | + | **5.Required reagents:** |
● R1. α-glucosidase 16.000 KU/L. | ● R1. α-glucosidase 16.000 KU/L. | ||
● R2. Ethylidene-4-NP-G7 (EPS) 6.501 g/L. | ● R2. Ethylidene-4-NP-G7 (EPS) 6.501 g/L. | ||
- | 5.Reagent Handling | + | **6.Reagent Handling** |
- | ● Remove any air bubbles present in the reagents with a new applicator stick, or allow the reagents to settle at the appropriate storage temperature to allow the bubbles to dissipate. To minimize volume depletion, do not use a transfer pipette to remove bubbles | + | |
- | **6.Reagent Storage and stability** | + | |
+ | **7.Reagent Storage and stability** | ||
Unopened reagent stable at 15-30°C until expiration date. | Unopened reagent stable at 15-30°C until expiration date. | ||
On board System temperature reagent is stable for 30 days | On board System temperature reagent is stable for 30 days | ||
Line 36: | Line 37: | ||
contamination, | contamination, | ||
- | 7.Calibration Procedure: | + | **8.Calibration Procedure:** |
- | ● Consolidated Chemistry | + | |
- | ● Frequency: | + | |
- | | + | |
- | | + | |
- | | + | |
- | | + | |
**Procedure: | **Procedure: | ||
Line 54: | Line 55: | ||
- If control results fall outside acceptable ranges,root cause analysis or recalibration may be necessary. | - If control results fall outside acceptable ranges,root cause analysis or recalibration may be necessary. | ||
- | **8.Quality control Procedure: | + | **9.Quality control Procedure: |
* Name: Bio Rad Level 1 &2 | * Name: Bio Rad Level 1 &2 | ||
* Frequency: As per Quality Control Procedure | * Frequency: As per Quality Control Procedure | ||
Line 75: | Line 76: | ||
- Check IQC results, in case outliers call residents. | - Check IQC results, in case outliers call residents. | ||
- | **9.Principle of the procedure used for examinations: | + | **10.Principle of the procedure used for examinations: |
At an alkaline pH, creatinine in the sample reacts with picrate to form a creatinine-picrate complex. The rate of increase in absorbance at 500 nm due to the formation of this complex is directly proportional to the concentration of creatinine in the sample. | At an alkaline pH, creatinine in the sample reacts with picrate to form a creatinine-picrate complex. The rate of increase in absorbance at 500 nm due to the formation of this complex is directly proportional to the concentration of creatinine in the sample. | ||
- | **10.Sample Preparation: | + | **11.Sample Preparation: |
* Required SampleVolume: | * Required SampleVolume: | ||
* Temperature: | * Temperature: | ||
Line 91: | Line 92: | ||
- Put respected carrier in RSH rack | - Put respected carrier in RSH rack | ||
- | **11.Performance Characteristics: | + | **12.Performance Characteristics: |
* Linearity: | * Linearity: | ||
* The limit of detection (LOD): 1 U/L | * The limit of detection (LOD): 1 U/L | ||
Line 101: | Line 102: | ||
^Adult^28-100^U/ | ^Adult^28-100^U/ | ||
- | **12.Laboratory Clinical interpretation: | + | **13.Laboratory Clinical interpretation: |
Amylase is used primarily in the diagnosis of acute pancreatitis. Acute pancreatitis is a reversible inflammation due to enzymatic necrosis. The most common cause of acute pancreatitis is gallstones. The second most common cause is heavy alcohol consumption. Serum amylase elevation and concomitant urine amylase elevation is seen in patients with acute pancreatitis. | Amylase is used primarily in the diagnosis of acute pancreatitis. Acute pancreatitis is a reversible inflammation due to enzymatic necrosis. The most common cause of acute pancreatitis is gallstones. The second most common cause is heavy alcohol consumption. Serum amylase elevation and concomitant urine amylase elevation is seen in patients with acute pancreatitis. | ||
- | **13.Interference and cross reaction:** | + | **14.Interference and cross reaction:** |
The Following analytes were tested up to the levels indicated at Creatinine concentrations of(approximately 50 U/L and 200 U/L). mg/dl, and found not to interfere: | The Following analytes were tested up to the levels indicated at Creatinine concentrations of(approximately 50 U/L and 200 U/L). mg/dl, and found not to interfere: | ||
Avoid contaminents like saliva, | Avoid contaminents like saliva, | ||
- | **14.Potential source of variation: | + | **15.Potential source of variation: |
- Turn around time (TAT): | - Turn around time (TAT): | ||
- Routine: 6.0 hours | - Routine: 6.0 hours | ||
- Urgent: 2.0 hours | - Urgent: 2.0 hours | ||
- | **15.Recording of observation: | + | **16.Recording of observation: |
* Software backup | * Software backup | ||
* Machine raw data | * Machine raw data | ||
- | **16.Storage & Disposal of waste:** Follow storage & discard procedure | + | **17.Storage & Disposal of waste:** Follow storage & discard procedure |
- | **17.Environmental & Safety control:** | + | **18.Environmental & Safety control:** |
Contact with acids liberates very toxic gas. Dispose of contents / container in accordance with local regulations. | Contact with acids liberates very toxic gas. Dispose of contents / container in accordance with local regulations. | ||
- | **18.References**: | + | **19.References**: |
- Tietz NW, Huang WY, Rauh DF, et al. Laboratory tests in the differential diagnosis of hyperamylasemia. Clin Chem 1986 32(2): | - Tietz NW, Huang WY, Rauh DF, et al. Laboratory tests in the differential diagnosis of hyperamylasemia. Clin Chem 1986 32(2): | ||
- 6 Junge W. Troge B, Klein G, et al. Evaluation of a New Assay for Pancreatic Amylase: Performance Characteristics and Estimation of Reference Intervals Clin Biochem 1989,22 109-114 | - 6 Junge W. Troge B, Klein G, et al. Evaluation of a New Assay for Pancreatic Amylase: Performance Characteristics and Estimation of Reference Intervals Clin Biochem 1989,22 109-114 |